This pilot study aimed to establish end-to-end stereotactic radiosurgery (SRS) sessions and evaluate the accuracy of the delivered doses using EPR/alanine and GAFChromic™ EBT3 films. Six Planning Treatment Volumes (PTVs), representing variations in tumor size, location, and proximity to Organs at Risk (OARs), were assessed using an anthropomorphic male head phantom irradiated with a TomoTherapy Hi-Art system (fraction of 10.5 Gy). Maximum deviations ranged from − 0.35 to 0.16 Gy, and − 0.35 to − 0.08 Gy, for alanine and films, respectively, even under complex spatial dose distribution (e.g., PTV6). Therefore, the findings demonstrated the reliability of both systems for precise SRS dosimetry.

Panzeca, S., De Farias Soares, A., Iacoviello, G., Cucchiara, T., Gallo, S., Romeo, M., et al. (2025). Dosimetric evaluation using EPR/alanine and radiochromic film detectors in end-to-end brain stereotactic radiosurgery sessions. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 334(10), 7305-7317 [10.1007/s10967-025-10429-z].

Dosimetric evaluation using EPR/alanine and radiochromic film detectors in end-to-end brain stereotactic radiosurgery sessions

de Farias Soares A.
;
Romeo M.;D'Oca M. C.;Marrale M.
2025-10-25

Abstract

This pilot study aimed to establish end-to-end stereotactic radiosurgery (SRS) sessions and evaluate the accuracy of the delivered doses using EPR/alanine and GAFChromic™ EBT3 films. Six Planning Treatment Volumes (PTVs), representing variations in tumor size, location, and proximity to Organs at Risk (OARs), were assessed using an anthropomorphic male head phantom irradiated with a TomoTherapy Hi-Art system (fraction of 10.5 Gy). Maximum deviations ranged from − 0.35 to 0.16 Gy, and − 0.35 to − 0.08 Gy, for alanine and films, respectively, even under complex spatial dose distribution (e.g., PTV6). Therefore, the findings demonstrated the reliability of both systems for precise SRS dosimetry.
25-ott-2025
Settore PHYS-06/A - Fisica per le scienze della vita, l'ambiente e i beni culturali
Panzeca, S., De Farias Soares, A., Iacoviello, G., Cucchiara, T., Gallo, S., Romeo, M., et al. (2025). Dosimetric evaluation using EPR/alanine and radiochromic film detectors in end-to-end brain stereotactic radiosurgery sessions. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 334(10), 7305-7317 [10.1007/s10967-025-10429-z].
File in questo prodotto:
File Dimensione Formato  
Dosimetric evaluation using EPR_alanine and radiochromic film.pdf

accesso aperto

Descrizione: articolo
Tipologia: Versione Editoriale
Dimensione 2.24 MB
Formato Adobe PDF
2.24 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/700903
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact