A full-energy-peak efficiency (FEPE) calibration procedure for gamma-ray spectrometric analysis of air particulate samples collected on large filters is described herein. The experimental results are obtained for an unconventional measurement geometry, termed a "packet-sample". The sample is obtained from a large cellulose filter (45 cm × 45 cm) used to collect air particulate samples that is resized to dimensions suitable for spectrometric measurements (6 cm×6 cm×0.7 cm). To determine the FEPEs, many standards were created, i.e., some filters containing a small amount of ThO2 and others containing a known amount of KCl. Efficiency curves obtained through best fits to experimental data for three high-purity germanium (HPGe) detectors, one of them for X-ray measurement, cover the energy range from 40 keV to 2600 keV. To validate the efficiency calibration procedure two experimental tests with the use of calibrated sources were conducted and, as application example, airborne concentrations of 131I (particulate matter), 134Cs and 137Cs at Palermo (Italy) in the days following the Fukushima accident in Japan were evaluated.
Tomarchio, E. (2013). An experimental approach to efficiency calibration for gamma-ray spectrometric analysis of large air particulate filters. RADIATION PHYSICS AND CHEMISTRY, 53(85 (2013)), 53-58 [10.1016/j.radphyschem.2012.11.005].
An experimental approach to efficiency calibration for gamma-ray spectrometric analysis of large air particulate filters
TOMARCHIO, Elio Angelo
2013-01-01
Abstract
A full-energy-peak efficiency (FEPE) calibration procedure for gamma-ray spectrometric analysis of air particulate samples collected on large filters is described herein. The experimental results are obtained for an unconventional measurement geometry, termed a "packet-sample". The sample is obtained from a large cellulose filter (45 cm × 45 cm) used to collect air particulate samples that is resized to dimensions suitable for spectrometric measurements (6 cm×6 cm×0.7 cm). To determine the FEPEs, many standards were created, i.e., some filters containing a small amount of ThO2 and others containing a known amount of KCl. Efficiency curves obtained through best fits to experimental data for three high-purity germanium (HPGe) detectors, one of them for X-ray measurement, cover the energy range from 40 keV to 2600 keV. To validate the efficiency calibration procedure two experimental tests with the use of calibrated sources were conducted and, as application example, airborne concentrations of 131I (particulate matter), 134Cs and 137Cs at Palermo (Italy) in the days following the Fukushima accident in Japan were evaluated.File | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S0969806X12004902-main.pdf
Solo gestori archvio
Descrizione: File PDF
Dimensione
525.42 kB
Formato
Adobe PDF
|
525.42 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.