Tissue-engineered heart valve (TEHV) scaffolds are designed as off-the-shelf alternatives, eliminating the need for organ donors and promoting native tissue regeneration. This study presents an in silico investigation of a bioinspired, polymeric, stentless mitral valve (MV) prosthesis, focusing on engineering the chordal apparatus (CA) to optimize valve performance. Specifically, the study explores the influence of chordae tendineae (CT) number, length, and anchoring position on functional outcomes. Numerical simulations were conducted using Abaqus/Explicit. The mitral leaflets and CT were modelled using shell (S4) and truss (T3D2) elements, respectively. To replicate the anisotropic mechanical behaviour of native mitral leaflets, an anisotropic Fan–Sacks constitutive model was employed. The CTs were described with an Ogden hyperelastic model. Boundary conditions included an encastred annulus and papillary muscles (PM) pinned at their time-averaged positions. The leaflets were loaded with physiological atrial and ventricular pressures. Parametric variations included CT number (2, 4, 6, 8), CT length (80%, 100%, 120% of physiological length), and anchoring distribution along the leaflet free edge. Performance metrics evaluated included von Mises stress (σ ), bulging height (BH), coaptation length (CL), and geometric orifice area (GOA), benchmarked against native valve thresholds. The optimal configuration featured 8 CT (4 per leaflet), at 80% of physiological length, with anchoring points along the free edge. This setup enhanced CL (>5 mm) and GOA (>1.5 cm²), while reducing σ (<1.5 MPa) and BH (<2 mm). These results provide crucial guidance for the design of the engineered BIOMITRAL valve incorporating a bioinspired CA.

Lanzalaco, E.; Dario Soto Laubrie, J.; Cosentino, F.; Raffa, G.; Pilato, M.; La Carrubba, V.; Pantano, A.; D'Amore, A. (17-19 settembre).Toward a Bioinspired Tissue Engineered Mitral Valve: In Silico Optimization of the Chordal Apparatus in a Polymeric Mitral Valve.

Toward a Bioinspired Tissue Engineered Mitral Valve: In Silico Optimization of the Chordal Apparatus in a Polymeric Mitral Valve

Elisa Lanzalaco
Primo
;
Federica Cosentino;Giuseppe Raffa;Vincenzo La Carrubba;Antonio Pantano
Penultimo
;
Antonio D'Amore
Ultimo

Abstract

Tissue-engineered heart valve (TEHV) scaffolds are designed as off-the-shelf alternatives, eliminating the need for organ donors and promoting native tissue regeneration. This study presents an in silico investigation of a bioinspired, polymeric, stentless mitral valve (MV) prosthesis, focusing on engineering the chordal apparatus (CA) to optimize valve performance. Specifically, the study explores the influence of chordae tendineae (CT) number, length, and anchoring position on functional outcomes. Numerical simulations were conducted using Abaqus/Explicit. The mitral leaflets and CT were modelled using shell (S4) and truss (T3D2) elements, respectively. To replicate the anisotropic mechanical behaviour of native mitral leaflets, an anisotropic Fan–Sacks constitutive model was employed. The CTs were described with an Ogden hyperelastic model. Boundary conditions included an encastred annulus and papillary muscles (PM) pinned at their time-averaged positions. The leaflets were loaded with physiological atrial and ventricular pressures. Parametric variations included CT number (2, 4, 6, 8), CT length (80%, 100%, 120% of physiological length), and anchoring distribution along the leaflet free edge. Performance metrics evaluated included von Mises stress (σ ), bulging height (BH), coaptation length (CL), and geometric orifice area (GOA), benchmarked against native valve thresholds. The optimal configuration featured 8 CT (4 per leaflet), at 80% of physiological length, with anchoring points along the free edge. This setup enhanced CL (>5 mm) and GOA (>1.5 cm²), while reducing σ (<1.5 MPa) and BH (<2 mm). These results provide crucial guidance for the design of the engineered BIOMITRAL valve incorporating a bioinspired CA.
Mitral vave, numerical analysis, FEM, Abaqus, chordae tendineae distribution
Lanzalaco, E.; Dario Soto Laubrie, J.; Cosentino, F.; Raffa, G.; Pilato, M.; La Carrubba, V.; Pantano, A.; D'Amore, A. (17-19 settembre).Toward a Bioinspired Tissue Engineered Mitral Valve: In Silico Optimization of the Chordal Apparatus in a Polymeric Mitral Valve.
File in questo prodotto:
File Dimensione Formato  
Abstract_HVBTE2025_EL.pdf

accesso aperto

Tipologia: Versione Editoriale
Dimensione 67.51 kB
Formato Adobe PDF
67.51 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/692243
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact