The long-term effectiveness of osseointegrated implants is heavily dependent on the short-term stability, primarily achieved immediately after surgery through a mechanical connection between the bone and the implant. The most common implant designs nowadays are straight and rely on screw or press-fit fixtures. Despite the promising results achieved by current transfemoral implants, the incidence of early failures and complications is still high. Starting from the hypothesis that a patient-specific approach could lead to better primary stability immediately post-surgery, this study aims to investigate the effect of implant design on primary stability. This was performed by analyzing two patient-specific implants, customized according to the medullary canal morphology, and a simple straight implant as the reference standard. To quantitatively assess the primary stability, a comparative computational analysis was conducted to examine the effective contact area, the relative micromotion, and the stress distribution at the interface between the bone and the implant stem during a static load-bearing exercise. The results showed that implants that follow the curvature of the residual femur provide lower micromotion values and a wider contact area, with a reduction of up to 30.4% and an increase of 10.8%, respectively, compared to the straight design, leading to a more homogeneous load distribution. Patient-specific prosthetic implants allow a more homogenous contact distribution that could lead to higher primary stability by reducing micromotion at the bone-implant interface concerning the straight profile, lowering the risk of loosening related to the short-term stability.

Valenti A., Cirello A., Ingrassia T., Marannano G., Nigrelli V., Ricotta V., et al. (2025). Primary Stability Assessment of Patient-Specific Transfemoral Osseointegrated Implants During Static Load-Bearing Exercise: A Comparative Analysis. APPLIED SCIENCES, 15(11) [10.3390/app15115913].

Primary Stability Assessment of Patient-Specific Transfemoral Osseointegrated Implants During Static Load-Bearing Exercise: A Comparative Analysis

Valenti A.
Primo
;
Cirello A.;Ingrassia T.;Marannano G.;Nigrelli V.;Ricotta V.;Mirulla A. I.
Ultimo
2025-06-01

Abstract

The long-term effectiveness of osseointegrated implants is heavily dependent on the short-term stability, primarily achieved immediately after surgery through a mechanical connection between the bone and the implant. The most common implant designs nowadays are straight and rely on screw or press-fit fixtures. Despite the promising results achieved by current transfemoral implants, the incidence of early failures and complications is still high. Starting from the hypothesis that a patient-specific approach could lead to better primary stability immediately post-surgery, this study aims to investigate the effect of implant design on primary stability. This was performed by analyzing two patient-specific implants, customized according to the medullary canal morphology, and a simple straight implant as the reference standard. To quantitatively assess the primary stability, a comparative computational analysis was conducted to examine the effective contact area, the relative micromotion, and the stress distribution at the interface between the bone and the implant stem during a static load-bearing exercise. The results showed that implants that follow the curvature of the residual femur provide lower micromotion values and a wider contact area, with a reduction of up to 30.4% and an increase of 10.8%, respectively, compared to the straight design, leading to a more homogeneous load distribution. Patient-specific prosthetic implants allow a more homogenous contact distribution that could lead to higher primary stability by reducing micromotion at the bone-implant interface concerning the straight profile, lowering the risk of loosening related to the short-term stability.
giu-2025
Settore IIND-03/B - Disegno e metodi dell'ingegneria industriale
Valenti A., Cirello A., Ingrassia T., Marannano G., Nigrelli V., Ricotta V., et al. (2025). Primary Stability Assessment of Patient-Specific Transfemoral Osseointegrated Implants During Static Load-Bearing Exercise: A Comparative Analysis. APPLIED SCIENCES, 15(11) [10.3390/app15115913].
File in questo prodotto:
File Dimensione Formato  
applsci-15-05913 (1).pdf

accesso aperto

Descrizione: This article is an open access article distributed under the termsand conditions of the Creative Commons Attribution (CC BY)license (https://creativecommons.org/ licenses/by/4.0/).
Tipologia: Versione Editoriale
Dimensione 3.05 MB
Formato Adobe PDF
3.05 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/683645
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact