The neutrino mass experiment KATRIN requires a stability of 3 ppm for the retarding potential at − 18.6 kV of the main spectrometer. To monitor the stability, two custom-made ultra-precise high-voltage dividers were developed and built in cooperation with the German national metrology institute Physikalisch-Technische Bundesanstalt (PTB). Until now, regular absolute calibration of the voltage dividers required bringing the equipment to the specialised metrology laboratory. Here we present a new method based on measuring the energy difference of two Kr conversion electron lines with the KATRIN setup, which was demonstrated during KATRIN’s commissioning measurements in July 2017. The measured scale factor of the high-voltage divider K35 is in agreement with the last PTB calibration 4 years ago. This result demonstrates the utility of the calibration method, as well as the long-term stability of the voltage divider.

Arenz M, Baek WJ, Beck M, Beglarian A, Behrens J, Bergmann T, et al. (2018). Calibration of high voltages at the ppm level by the difference of Kr-83m conversion electron lines at the KATRIN experiment. THE EUROPEAN PHYSICAL JOURNAL. C, PARTICLES AND FIELDS, 78(5) [10.1140/epjc/s10052-018-5832-y].

Calibration of high voltages at the ppm level by the difference of Kr-83m conversion electron lines at the KATRIN experiment

Roccati F;
2018-05-09

Abstract

The neutrino mass experiment KATRIN requires a stability of 3 ppm for the retarding potential at − 18.6 kV of the main spectrometer. To monitor the stability, two custom-made ultra-precise high-voltage dividers were developed and built in cooperation with the German national metrology institute Physikalisch-Technische Bundesanstalt (PTB). Until now, regular absolute calibration of the voltage dividers required bringing the equipment to the specialised metrology laboratory. Here we present a new method based on measuring the energy difference of two Kr conversion electron lines with the KATRIN setup, which was demonstrated during KATRIN’s commissioning measurements in July 2017. The measured scale factor of the high-voltage divider K35 is in agreement with the last PTB calibration 4 years ago. This result demonstrates the utility of the calibration method, as well as the long-term stability of the voltage divider.
9-mag-2018
Settore PHYS-01/A - Fisica sperimentale delle interazioni fondamentali e applicazioni
Arenz M, Baek WJ, Beck M, Beglarian A, Behrens J, Bergmann T, et al. (2018). Calibration of high voltages at the ppm level by the difference of Kr-83m conversion electron lines at the KATRIN experiment. THE EUROPEAN PHYSICAL JOURNAL. C, PARTICLES AND FIELDS, 78(5) [10.1140/epjc/s10052-018-5832-y].
File in questo prodotto:
File Dimensione Formato  
2018_Article_.pdf

accesso aperto

Descrizione: articolo
Tipologia: Versione Editoriale
Dimensione 1.25 MB
Formato Adobe PDF
1.25 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/676347
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 23
  • ???jsp.display-item.citation.isi??? 18
social impact