The study aim was to assess, in vitro, the hemodynamic modifications produced by transcatheter valves in the Valsalva sinuses, by mean of phase-resolved particle image velocimetry (PIV) measurements. Flow measurements were performed on a glass mock aortic root that included three polymeric valve leaflets, before and after the implantation of a Medtronic CoreValve device and of an Edwards SAPIEN valve. All experiments were carried out in a hydro-mechanical cardiovascular pulse duplicator system (Vivitro Superpump System SP3891) that reproduced physiologically equivalent pressures and flow rates conforming to the requirements of the standard ISO 5840:2005. The flow dynamics, before and after implantation of the two prosthetic devices, was characterized on the basis of phase-resolved velocity field and viscous shear rate measurements. Direct comparison indicated that both transcatheter valves determined a significant variation of flow during the early stages of valve opening and during valve closure. In general, the presence of the two valve implants significantly reduced the flow activity in the Valsalva sinuses, promoting regions of stagnation at their base. The reduction in flow in the Valsalva sinuses could be associated with the higher incidence of ischemic events reported after transcatheter heart valve implantation.

Ducci A., Tzamtzis S., Mullen M.J., Burriesci G. (2013). Hemodynamics in the Valsalva sinuses after transcatheter aortic valve implantation (TAVI). JOURNAL OF HEART VALVE DISEASE, 22(5), 688-696.

Hemodynamics in the Valsalva sinuses after transcatheter aortic valve implantation (TAVI)

Burriesci G.
Ultimo
2013-01-01

Abstract

The study aim was to assess, in vitro, the hemodynamic modifications produced by transcatheter valves in the Valsalva sinuses, by mean of phase-resolved particle image velocimetry (PIV) measurements. Flow measurements were performed on a glass mock aortic root that included three polymeric valve leaflets, before and after the implantation of a Medtronic CoreValve device and of an Edwards SAPIEN valve. All experiments were carried out in a hydro-mechanical cardiovascular pulse duplicator system (Vivitro Superpump System SP3891) that reproduced physiologically equivalent pressures and flow rates conforming to the requirements of the standard ISO 5840:2005. The flow dynamics, before and after implantation of the two prosthetic devices, was characterized on the basis of phase-resolved velocity field and viscous shear rate measurements. Direct comparison indicated that both transcatheter valves determined a significant variation of flow during the early stages of valve opening and during valve closure. In general, the presence of the two valve implants significantly reduced the flow activity in the Valsalva sinuses, promoting regions of stagnation at their base. The reduction in flow in the Valsalva sinuses could be associated with the higher incidence of ischemic events reported after transcatheter heart valve implantation.
2013
Ducci A., Tzamtzis S., Mullen M.J., Burriesci G. (2013). Hemodynamics in the Valsalva sinuses after transcatheter aortic valve implantation (TAVI). JOURNAL OF HEART VALVE DISEASE, 22(5), 688-696.
File in questo prodotto:
File Dimensione Formato  
Ducci 2013.pdf

Solo gestori archvio

Tipologia: Versione Editoriale
Dimensione 851.98 kB
Formato Adobe PDF
851.98 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/667569
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 51
  • ???jsp.display-item.citation.isi??? 50
social impact