The detection of serum anti-acetylcholine receptor (AChR) antibodies is currently an important tool for diagnosing myasthenia gravis (MG) since they are present in about 85% of MG patients. Many serological tests are now available. Nevertheless, results from these tests can be different in some patients. The aim of this study is to compare the sensitivity of a commercially available fixed cell-based assay (F-CBA) to that of enzyme-linked immunosorbent assay (ELISA) kits for anti-AChR detection in patients with a diagnosis of MG. Overall, 143 patients with a confirmed MG diagnosis were included in the study. The detection and measurement of serum anti-AChR antibodies were performed by three analytical methods, namely, a competitive ELISA (cELISA), an indirect ELISA (iELISA), and an F-CBA, according to the manufacturers' instructions. Anti-AChR antibody titers were positive in 94/143 (66%) using the cELISA, in 75/143 (52%) using the iELISA and in 61/143 (43%) using the F-CBA (adult and/or fetal). Method agreement, evaluated by concordant pairs and Cohen's kappa, was as follows: cELISA-iELISA: 110/143 (77%), k = 0.53 (95%CI 0.40-0.66); cELISA-F-CBA: 108/143 (76%), k = 0.53 (95%CI 0.41-0.66); iELISA-F-CBA: 121/143 (85%), k = 0.70 (95%CI 0.57-0.80). Our findings show that the cELISA has better analytical performance than the iELISA and F-CBA. However, the iELISA and F-CBA show the highest concordance.
Gambino, C.M., Agnello, L., Ciaccio, A.M., Scazzone, C., Vidali, M., Di Stefano, V., et al. (2023). Detection of Antibodies against the Acetylcholine Receptor in Patients with Myasthenia Gravis: A Comparison of Two Enzyme Immunoassays and a Fixed Cell-Based Assay. JOURNAL OF CLINICAL MEDICINE, 12(14), 1-10 [10.3390/jcm12144781].
Detection of Antibodies against the Acetylcholine Receptor in Patients with Myasthenia Gravis: A Comparison of Two Enzyme Immunoassays and a Fixed Cell-Based Assay
Gambino, Caterina Maria;Agnello, Luisa;Ciaccio, Anna Maria;Scazzone, Concetta;Di Stefano, Vincenzo;Milano, Salvatore;Brighina, Filippo;Candore, Giuseppina;Lo Sasso, Bruna;Ciaccio, Marcello
2023-07-19
Abstract
The detection of serum anti-acetylcholine receptor (AChR) antibodies is currently an important tool for diagnosing myasthenia gravis (MG) since they are present in about 85% of MG patients. Many serological tests are now available. Nevertheless, results from these tests can be different in some patients. The aim of this study is to compare the sensitivity of a commercially available fixed cell-based assay (F-CBA) to that of enzyme-linked immunosorbent assay (ELISA) kits for anti-AChR detection in patients with a diagnosis of MG. Overall, 143 patients with a confirmed MG diagnosis were included in the study. The detection and measurement of serum anti-AChR antibodies were performed by three analytical methods, namely, a competitive ELISA (cELISA), an indirect ELISA (iELISA), and an F-CBA, according to the manufacturers' instructions. Anti-AChR antibody titers were positive in 94/143 (66%) using the cELISA, in 75/143 (52%) using the iELISA and in 61/143 (43%) using the F-CBA (adult and/or fetal). Method agreement, evaluated by concordant pairs and Cohen's kappa, was as follows: cELISA-iELISA: 110/143 (77%), k = 0.53 (95%CI 0.40-0.66); cELISA-F-CBA: 108/143 (76%), k = 0.53 (95%CI 0.41-0.66); iELISA-F-CBA: 121/143 (85%), k = 0.70 (95%CI 0.57-0.80). Our findings show that the cELISA has better analytical performance than the iELISA and F-CBA. However, the iELISA and F-CBA show the highest concordance.File | Dimensione | Formato | |
---|---|---|---|
jcm-12-04781.pdf
accesso aperto
Tipologia:
Versione Editoriale
Dimensione
1.48 MB
Formato
Adobe PDF
|
1.48 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.