Simplified expressions to determine the coincidence-summing correction factors by using a matrix representation of a decay scheme are reported. Expressions can be used directly for point sources, with the introduction of full-energy-peak and total efficiencies, while for extended sources a more complex procedure, though approximate, is suggested.
Tomarchio, E.A. (2010). Coincidence-Summing corrections in high resolution gamma-ray spectrometry: simplified analytical expressions. In Atti Workshop "Metodi avanzati di spettrometria gamma".
Coincidence-Summing corrections in high resolution gamma-ray spectrometry: simplified analytical expressions
TOMARCHIO, Elio Angelo
2010-01-01
Abstract
Simplified expressions to determine the coincidence-summing correction factors by using a matrix representation of a decay scheme are reported. Expressions can be used directly for point sources, with the introduction of full-energy-peak and total efficiencies, while for extended sources a more complex procedure, though approximate, is suggested.File in questo prodotto: