Background: The detection of anti-acetylcholine receptor (AChR) and anti-muscle-specific tyrosine kinase (MuSK) antibodies is useful in myasthenia gravis (MG) diagnosis and management. BIOCHIP mosaic-based indirect immunofluorescence is a novel analytical method, which employs the simultaneous detection of anti-AChR and anti-MuSK antibodies in a single miniature incubation field. In this study, we compare, for the first time, the BIOCHIP MG mosaic with conventional enzyme-linked immunosorbent assay (ELISA) in the diagnosis of MG. Methods: A total of 71 patients with MG diagnosis were included in the study. Anti-AChR and anti-MuSK antibodies were measured separately by two different ELISA and simultaneously by BIOCHIP. The results were then compared. Results: The overall concordance between ELISA and BIOCHIP for anti-AChR reactivity was 74%. Cohen’s kappa was 0.51 (95% CI 0.32–0.71), which corresponds to 90% of the maximum possible kappa (0.57), given the observed marginal frequencies. The overall concordance for anti-MuSK reactivity was 84%. Cohen’s kappa was 0.11 (95% CI 0.00–0.36), which corresponds to 41% of the maximum possible kappa (0.27). Conclusion: The overall concordance among assays is not optimal.
Gambino C.M., Agnello L., Lo Sasso B., Scazzone C., Giglio R.V., Candore G., et al. (2021). Comparative analysis of biochip mosaic-based indirect immunofluorescence with enzyme-linked immunosorbent assay for diagnosing myasthenia gravis. DIAGNOSTICS, 11(11), 2098 [10.3390/diagnostics11112098].
Comparative analysis of biochip mosaic-based indirect immunofluorescence with enzyme-linked immunosorbent assay for diagnosing myasthenia gravis
Gambino C. M.;Agnello L.;Lo Sasso B.;Scazzone C.;Giglio R. V.;Candore G.;Ciaccio A. M.;Di Stefano V.;Brighina F.;Ciaccio M.
2021-11-13
Abstract
Background: The detection of anti-acetylcholine receptor (AChR) and anti-muscle-specific tyrosine kinase (MuSK) antibodies is useful in myasthenia gravis (MG) diagnosis and management. BIOCHIP mosaic-based indirect immunofluorescence is a novel analytical method, which employs the simultaneous detection of anti-AChR and anti-MuSK antibodies in a single miniature incubation field. In this study, we compare, for the first time, the BIOCHIP MG mosaic with conventional enzyme-linked immunosorbent assay (ELISA) in the diagnosis of MG. Methods: A total of 71 patients with MG diagnosis were included in the study. Anti-AChR and anti-MuSK antibodies were measured separately by two different ELISA and simultaneously by BIOCHIP. The results were then compared. Results: The overall concordance between ELISA and BIOCHIP for anti-AChR reactivity was 74%. Cohen’s kappa was 0.51 (95% CI 0.32–0.71), which corresponds to 90% of the maximum possible kappa (0.57), given the observed marginal frequencies. The overall concordance for anti-MuSK reactivity was 84%. Cohen’s kappa was 0.11 (95% CI 0.00–0.36), which corresponds to 41% of the maximum possible kappa (0.27). Conclusion: The overall concordance among assays is not optimal.File | Dimensione | Formato | |
---|---|---|---|
Comparative Analysis of BIOCHIP Mosaic-Based Indirect Immunofluorescence with Enzyme-Linked Immunosorbent Assay for Diagnosing Myasthenia Gravis.pdf