The high performances of CdTe detectors for X-ray and gamma ray spectroscopy are already well known. Among the traditional semiconductor spectrometers, CdTe detectors show high detection efficiency and good room temperature performance and are well suited for the development of compact detection systems. In this work, we investigated the performance of a CdTe detector coupled with a custom digital pulse processing (DPP) system for X-ray spectroscopy. The DPP method, implemented on a PC platform, performs a pile-up inspection and a pulse height analysis of the preamplifier output pulses, digitized by a 14-bit, 100 MHz ADC. The spectroscopic results point out the excellent performance of the digital spectrometer both at low (150 cps) and at high photon counting rate (up to 370 kcps), confirming the high potentialities of CdTe detectors coupled with DPP systems. The detector and the DPP system were developed by our collaboration as a spectrometer prototype for direct measurement of diagnostic X-ray spectra in the 1–40 keV energy range
Abbene, L., Gerardi, G., Del Sordo, S., Raso, G. (2010). Performance of a digital CdTe X-ray spectrometer in low and high counting rate environment. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 621(1-3), 447-452 [10.1016/j.nima.2010.06.002].
Performance of a digital CdTe X-ray spectrometer in low and high counting rate environment
ABBENE, Leonardo;GERARDI, Gaetano;RASO, Giuseppe
2010-01-01
Abstract
The high performances of CdTe detectors for X-ray and gamma ray spectroscopy are already well known. Among the traditional semiconductor spectrometers, CdTe detectors show high detection efficiency and good room temperature performance and are well suited for the development of compact detection systems. In this work, we investigated the performance of a CdTe detector coupled with a custom digital pulse processing (DPP) system for X-ray spectroscopy. The DPP method, implemented on a PC platform, performs a pile-up inspection and a pulse height analysis of the preamplifier output pulses, digitized by a 14-bit, 100 MHz ADC. The spectroscopic results point out the excellent performance of the digital spectrometer both at low (150 cps) and at high photon counting rate (up to 370 kcps), confirming the high potentialities of CdTe detectors coupled with DPP systems. The detector and the DPP system were developed by our collaboration as a spectrometer prototype for direct measurement of diagnostic X-ray spectra in the 1–40 keV energy rangeFile | Dimensione | Formato | |
---|---|---|---|
L. Abbene_NIMA621_2010.pdf
Solo gestori archvio
Dimensione
851.24 kB
Formato
Adobe PDF
|
851.24 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.