X-ray spectra and dosimetry information are very important for quality assurance (QA) and quality control (QC) in medical diagnostic X-ray systems. An accurate knowledge of the diagnostic X-ray spectra would improve the patient dose optimization, without compromising image information. In this work, we performed direct diagnostic X-ray spectra measurements with a portable device, based on a CdTe solid-state detector. The portable device is able to directly measure X-ray spectra at high photon fluence rates, as typical of clinical radiography. We investigated on the spectral performances of the system in the mammographic energy range (up to ∼40 keV). Good system response to monoenergetic photons was measured (energy resolution of 5% FWHM at 22.1 keV). We measured the molybdenum X-ray spectra produced by a mammographic X-ray unit (GE Senographe DMR) at 28 kV and 30 kV under clinical conditions. The results showed the good reproducibility of the system and low pile-up distortions. Preliminary dosimetric measurements have been regarded as exposure and half value layer (HVL) values obtained from direct measurements and from measured X-ray spectral data, and a good agreement between exposure attenuation curves and the HVL values was obtained. The results indicated that the portable device is suitable for mammographic X-ray spectroscopy under clinical conditions

ABBENE L, LA MANNA A, FAUCI F, GERARDI G, STUMBO S, RASO G (2007). X-ray spectroscopy and dosimetry with a portable CdTe device. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 571(1-2), 373-377 [10.1016/j.nima.2006.10.112].

X-ray spectroscopy and dosimetry with a portable CdTe device.

ABBENE, Leonardo;FAUCI, Francesco;GERARDI, Gaetano;RASO, Giuseppe
2007-01-01

Abstract

X-ray spectra and dosimetry information are very important for quality assurance (QA) and quality control (QC) in medical diagnostic X-ray systems. An accurate knowledge of the diagnostic X-ray spectra would improve the patient dose optimization, without compromising image information. In this work, we performed direct diagnostic X-ray spectra measurements with a portable device, based on a CdTe solid-state detector. The portable device is able to directly measure X-ray spectra at high photon fluence rates, as typical of clinical radiography. We investigated on the spectral performances of the system in the mammographic energy range (up to ∼40 keV). Good system response to monoenergetic photons was measured (energy resolution of 5% FWHM at 22.1 keV). We measured the molybdenum X-ray spectra produced by a mammographic X-ray unit (GE Senographe DMR) at 28 kV and 30 kV under clinical conditions. The results showed the good reproducibility of the system and low pile-up distortions. Preliminary dosimetric measurements have been regarded as exposure and half value layer (HVL) values obtained from direct measurements and from measured X-ray spectral data, and a good agreement between exposure attenuation curves and the HVL values was obtained. The results indicated that the portable device is suitable for mammographic X-ray spectroscopy under clinical conditions
2007
ABBENE L, LA MANNA A, FAUCI F, GERARDI G, STUMBO S, RASO G (2007). X-ray spectroscopy and dosimetry with a portable CdTe device. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 571(1-2), 373-377 [10.1016/j.nima.2006.10.112].
File in questo prodotto:
File Dimensione Formato  
x-ray.pdf

Solo gestori archvio

Dimensione 141.06 kB
Formato Adobe PDF
141.06 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/4800
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 36
  • ???jsp.display-item.citation.isi??? 34
social impact