The study reported in this paper aimed to investigate the response of a GAFChromic (R) HD-V2 film dosimeter in the dose range between 10 and 300 Gy, normally used to perform irradiation tests on electronic devices with a Co-60 gamma-ray irradiator. The well-defined linearity of response in terms of absorbance as a function of absorbed dose, easiness of handling and data analysis of irradiated dosimeters, reproducibility, stability; and insensitivity to visible light and most of the environmental parameters, make HD-V2 film a flexible, inexpensive and reliable dose measurement device. The study has shown a fogging effect of the dosimeter response since its value changes over time. Strategies and adequate precautions to take into account changes on HD-V2 film physical properties for calibration and routine use in industrial applications, with a specific reference to radiation testing of electronic devices, have been discussed.
Martorana F, Parlato A, Perrone G, Tomarchio E (2019). Response Of Gafchromic® Hd-V2 Film Dosimeter In 10-300 Gy Dose Range For Radiation Testing Of Electronic Devices. NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 34(1), 79-84 [10.2298/NTRP180801005M].
Response Of Gafchromic® Hd-V2 Film Dosimeter In 10-300 Gy Dose Range For Radiation Testing Of Electronic Devices
Martorana F;Parlato A;Tomarchio E
2019-01-01
Abstract
The study reported in this paper aimed to investigate the response of a GAFChromic (R) HD-V2 film dosimeter in the dose range between 10 and 300 Gy, normally used to perform irradiation tests on electronic devices with a Co-60 gamma-ray irradiator. The well-defined linearity of response in terms of absorbance as a function of absorbed dose, easiness of handling and data analysis of irradiated dosimeters, reproducibility, stability; and insensitivity to visible light and most of the environmental parameters, make HD-V2 film a flexible, inexpensive and reliable dose measurement device. The study has shown a fogging effect of the dosimeter response since its value changes over time. Strategies and adequate precautions to take into account changes on HD-V2 film physical properties for calibration and routine use in industrial applications, with a specific reference to radiation testing of electronic devices, have been discussed.File | Dimensione | Formato | |
---|---|---|---|
Tomarchio_2019_1.pdf
accesso aperto
Descrizione: articolo su rivista ISI
Tipologia:
Versione Editoriale
Dimensione
471.67 kB
Formato
Adobe PDF
|
471.67 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.