This work deals with the dosimetric features of a particular phenolic compound (IRGANOX 1076 ® ) for dosimetry of clinical photon beams by using electron spin resonance (ESR) spectroscopy. After the optimization of the ESR readout parameters (namely modulation amplitude and microwave power) to maximise the signal without excessive spectrum distortions, basic dosimetric properties of laboratory-made phenolic dosimeters in pellet form, such as reproducibility, dose–response, sensitivity, linearity and dose rate dependence were investigated. The dosimeters were tested by measuring the depth dose profile of a 6 MV photon beam. A satisfactory intra-batch reproducibility of the ESR signal of the manufactured dosimeters was obtained. The ESR signal proved to increase linearly with increasing dose in the investigated dose range 1–13 Gy. The presence of an intrinsic background signal limits the minimum detectable dose to a value of approximately 0.6 Gy. Reliable and accurate assessment of the dose was achieved, independently of the dose rate. Such characteristics, together with the fact that IRGANOX 1076 ® is almost tissue-equivalent, and the stability of the ESR signal, make these dosimeters promising materials for ESR dosimetric applications in radiotherapy.

Gallo S., Iacoviello G., Panzeca S., Veronese I., Bartolotta A., Dondi D., et al. (2017). Characterization of phenolic pellets for ESR dosimetry in photon beam radiotherapy. RADIATION AND ENVIRONMENTAL BIOPHYSICS, 56(4), 471-480 [10.1007/s00411-017-0716-3].

Characterization of phenolic pellets for ESR dosimetry in photon beam radiotherapy

Panzeca S.;Marrale M.
2017-01-01

Abstract

This work deals with the dosimetric features of a particular phenolic compound (IRGANOX 1076 ® ) for dosimetry of clinical photon beams by using electron spin resonance (ESR) spectroscopy. After the optimization of the ESR readout parameters (namely modulation amplitude and microwave power) to maximise the signal without excessive spectrum distortions, basic dosimetric properties of laboratory-made phenolic dosimeters in pellet form, such as reproducibility, dose–response, sensitivity, linearity and dose rate dependence were investigated. The dosimeters were tested by measuring the depth dose profile of a 6 MV photon beam. A satisfactory intra-batch reproducibility of the ESR signal of the manufactured dosimeters was obtained. The ESR signal proved to increase linearly with increasing dose in the investigated dose range 1–13 Gy. The presence of an intrinsic background signal limits the minimum detectable dose to a value of approximately 0.6 Gy. Reliable and accurate assessment of the dose was achieved, independently of the dose rate. Such characteristics, together with the fact that IRGANOX 1076 ® is almost tissue-equivalent, and the stability of the ESR signal, make these dosimeters promising materials for ESR dosimetric applications in radiotherapy.
2017
Gallo S., Iacoviello G., Panzeca S., Veronese I., Bartolotta A., Dondi D., et al. (2017). Characterization of phenolic pellets for ESR dosimetry in photon beam radiotherapy. RADIATION AND ENVIRONMENTAL BIOPHYSICS, 56(4), 471-480 [10.1007/s00411-017-0716-3].
File in questo prodotto:
File Dimensione Formato  
2017_AA62_phenols_clinical_applications.pdf

Solo gestori archvio

Tipologia: Versione Editoriale
Dimensione 2.77 MB
Formato Adobe PDF
2.77 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Characterization of phenolic pellets for ESR dosimetry in photon beam radiotherapy..pdf

Solo gestori archvio

Descrizione: Articolo principale
Tipologia: Versione Editoriale
Dimensione 933.17 kB
Formato Adobe PDF
933.17 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/350829
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 17
social impact