NiO thin film thermally grown on Ni was investigated for the first time as an electrochemical sensor of mercury ions in water. The film was obtained by thermal oxidation in air of a commercial Ni foil. The influence of oxidation temperature, and time have been investigated by XRD, SEM, and EDS analysis. Square wave anodic stripping voltammetry was used for detecting Hg2+ions in aqueous solution with different ion concentrations. In order to improve the response signal, all the operational parameters related to the sensing process such as pH, deposition time, potential, and square wave frequency, have been optimized. We obtained a detection limit of 4.4 ppb with a sensitivity of 1.1 μA ppb-1cm-2.

Patella, B., Piazza, S., Sunseri, C., Inguanta, R. (2017). NiO thin film for mercury detection in water by square wave anodic stripping voltammetry. CHEMICAL ENGINEERING TRANSACTIONS, 60(60), 1-6 [10.3303/CET1760001].

NiO thin film for mercury detection in water by square wave anodic stripping voltammetry

Patella, Bernardo
Membro del Collaboration Group
;
Sunseri, Carmelo
Membro del Collaboration Group
;
Inguanta, Rosalinda
Membro del Collaboration Group
2017-01-01

Abstract

NiO thin film thermally grown on Ni was investigated for the first time as an electrochemical sensor of mercury ions in water. The film was obtained by thermal oxidation in air of a commercial Ni foil. The influence of oxidation temperature, and time have been investigated by XRD, SEM, and EDS analysis. Square wave anodic stripping voltammetry was used for detecting Hg2+ions in aqueous solution with different ion concentrations. In order to improve the response signal, all the operational parameters related to the sensing process such as pH, deposition time, potential, and square wave frequency, have been optimized. We obtained a detection limit of 4.4 ppb with a sensitivity of 1.1 μA ppb-1cm-2.
2017
Settore ING-IND/23 - Chimica Fisica Applicata
Patella, B., Piazza, S., Sunseri, C., Inguanta, R. (2017). NiO thin film for mercury detection in water by square wave anodic stripping voltammetry. CHEMICAL ENGINEERING TRANSACTIONS, 60(60), 1-6 [10.3303/CET1760001].
File in questo prodotto:
File Dimensione Formato  
CET-001-Inguanta-2017.pdf

Solo gestori archvio

Dimensione 893.86 kB
Formato Adobe PDF
893.86 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/252613
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 15
  • ???jsp.display-item.citation.isi??? ND
social impact