This paper presents a new technique to improve the quality of the ECG signals, increasing the Common Mode Rejection Ratio (CMRR). We developed a portable wireless Bluetooth ECG system able to acquire 12 leads, communicating with Smartphones and PCs. Many experiments have been made for measuring CMRR decay due to the difference on skin-electrodes impedances, the asymmetries of the amplifiers input stages and external components. Using some digital potentiometers, the system is able to compensate for these, increasing the CMRR of about 18 dB.
Di Buono P., Mistretta L., Giaconia, G. (2017). A novel technique for the CMRR improvement in a portable ECG system. In De Gloria A. (a cura di), Lecture Notes in Electrical Engineering (pp. 219-225). cham : Springer Verlag [10.1007/978-3-319-47913-2_25].
A novel technique for the CMRR improvement in a portable ECG system
DI BUONO, Pietro
;MISTRETTA, Leonardo;GIACONIA, Giuseppe Costantino
2017-01-01
Abstract
This paper presents a new technique to improve the quality of the ECG signals, increasing the Common Mode Rejection Ratio (CMRR). We developed a portable wireless Bluetooth ECG system able to acquire 12 leads, communicating with Smartphones and PCs. Many experiments have been made for measuring CMRR decay due to the difference on skin-electrodes impedances, the asymmetries of the amplifiers input stages and external components. Using some digital potentiometers, the system is able to compensate for these, increasing the CMRR of about 18 dB.File | Dimensione | Formato | |
---|---|---|---|
Giaconia.pdf
Solo gestori archvio
Descrizione: pdf
Tipologia:
Versione Editoriale
Dimensione
508.2 kB
Formato
Adobe PDF
|
508.2 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.