PURPOSE We aimed to compare the outcomes of the Endurant II (Medtronic) stent-graft used under instructions for use versus off-label in high-risk patients considered unfit for conventional surgery. METHODS Data from patients treated with the Endurant II stent-graft between December 2012 and March 2015 were retrospectively analyzed. Sixty-four patients were included. Patients were assigned to group A if treated under instructions for use (n=34, 53%) and to group B if treated off-label (n=30, 47%). Outcome measures included perioperative mortality and morbidity, survival, freedom from reintervention, endoleak incidence, in-hospital length of stay, and mean stent-graft component used. Mean follow-up was 22.61±12 months (median, 21.06 months; range, 0–43 months). RESULTS One perioperative mortality (1.6%) and one perioperative complication (1.6%) occurred in group B. At two months follow-up, one iliac limb occlusion (1.6%) occurred in group A. No type I/III endoleaks were recorded. A type II endoleak was identified in three cases (4.7%). Overall survival at three years was 89% (97% for group A, 82% for group B; P = 0.428). Reintervention-free survival at three years was 97% for both groups (P = 0.991). A longer in-hospital stay was observed in group B (P = 0.012). CONCLUSION The Endurant II (Medtronic) new generation device was safe in off-label setting at mid-term follow-up. The off-label use of the Endurant II (Medtronic) is justified in patients considered unfit for conventional surgery. Larger studies are required in this subgroup of patients.
Pecoraro, F., Corte, G., Dinoto, E., Badalamenti, G., Bruno, S., Bajardi, G. (2016). Clinical outcomes of endurant II stent-graft for infrarenal aortic aneurysm repair: Comparison of on-label versus off-label use. DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 22(5), 450-454 [10.5152/dir.2016.15418].
Clinical outcomes of endurant II stent-graft for infrarenal aortic aneurysm repair: Comparison of on-label versus off-label use
PECORARO, Felice
;BAJARDI, Guido
2016-01-01
Abstract
PURPOSE We aimed to compare the outcomes of the Endurant II (Medtronic) stent-graft used under instructions for use versus off-label in high-risk patients considered unfit for conventional surgery. METHODS Data from patients treated with the Endurant II stent-graft between December 2012 and March 2015 were retrospectively analyzed. Sixty-four patients were included. Patients were assigned to group A if treated under instructions for use (n=34, 53%) and to group B if treated off-label (n=30, 47%). Outcome measures included perioperative mortality and morbidity, survival, freedom from reintervention, endoleak incidence, in-hospital length of stay, and mean stent-graft component used. Mean follow-up was 22.61±12 months (median, 21.06 months; range, 0–43 months). RESULTS One perioperative mortality (1.6%) and one perioperative complication (1.6%) occurred in group B. At two months follow-up, one iliac limb occlusion (1.6%) occurred in group A. No type I/III endoleaks were recorded. A type II endoleak was identified in three cases (4.7%). Overall survival at three years was 89% (97% for group A, 82% for group B; P = 0.428). Reintervention-free survival at three years was 97% for both groups (P = 0.991). A longer in-hospital stay was observed in group B (P = 0.012). CONCLUSION The Endurant II (Medtronic) new generation device was safe in off-label setting at mid-term follow-up. The off-label use of the Endurant II (Medtronic) is justified in patients considered unfit for conventional surgery. Larger studies are required in this subgroup of patients.File | Dimensione | Formato | |
---|---|---|---|
72_2016_Endurant_IFUvsnonIFU_DIR.pdf
Solo gestori archvio
Dimensione
522.02 kB
Formato
Adobe PDF
|
522.02 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.