We described a modified technique for implanting a bridging stent-graft into an iliac branched device. A 79-year-old male who had received aortobiiliac synthetic graft surgery for a ruptured abdominal aortic aneurysm six months earlier was admitted to our unit for treatment of a left common iliac aneurysm involving the origin of the hypogastric artery. A standard technique was unsuccessful at implanting the bridging stent-graft, and therefore a wire-loop guidewire over the graft bifurcation was used to stabilize the contralateral sheaths and to complete the implantation. © Turkish Society of Radiology 2012.

Bracale, U., Narese, D., Vitale, G., Maglione, F., Conti, P., Bajardi, G. (2013). A wire-loop technique for implantation of an iliac branched device in a patient with previous surgery for a ruptured abdominal aortic aneurysm. DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 19(1), 73-75 [10.4261/1305-3825.DIR.5727-12.1].

A wire-loop technique for implantation of an iliac branched device in a patient with previous surgery for a ruptured abdominal aortic aneurysm

BAJARDI, Guido
2013-01-01

Abstract

We described a modified technique for implanting a bridging stent-graft into an iliac branched device. A 79-year-old male who had received aortobiiliac synthetic graft surgery for a ruptured abdominal aortic aneurysm six months earlier was admitted to our unit for treatment of a left common iliac aneurysm involving the origin of the hypogastric artery. A standard technique was unsuccessful at implanting the bridging stent-graft, and therefore a wire-loop guidewire over the graft bifurcation was used to stabilize the contralateral sheaths and to complete the implantation. © Turkish Society of Radiology 2012.
2013
Bracale, U., Narese, D., Vitale, G., Maglione, F., Conti, P., Bajardi, G. (2013). A wire-loop technique for implantation of an iliac branched device in a patient with previous surgery for a ruptured abdominal aortic aneurysm. DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 19(1), 73-75 [10.4261/1305-3825.DIR.5727-12.1].
File in questo prodotto:
File Dimensione Formato  
pdf_DIR_500.pdf

accesso aperto

Dimensione 237.59 kB
Formato Adobe PDF
237.59 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/152033
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
social impact