Abstract Al/CdTe/Pt detectors are very attractive devices for high-resolution X-ray spectroscopy, even though they suffer from bias-induced time instability (polarization). Polarization phenomena cause a progressive time-degradation of the spectroscopic performance of the detectors, due to hole trapping and detrapping from deep acceptor levels that directly control the electric field distribution. In this work we present experimental investigations on the electric field profile of planar Al/CdTe/Pt detectors by means of Pockels effect measurements. The time/temperature dependence of the electric field was investigated in a long time window (up to 10 h) and the correlation with the reverse current transients was also studied. Two energy levels (0.62 eV and 1.16 eV) of the deep hole traps were measured, in agreement with our previous results obtained through electrical and spectroscopic approaches.

Turturici, A., Abbene, L., Franc, J., Grill, R., Dědič, V., Principato, F. (2015). Time-dependent electric field in Al/CdTe/Pt detectors. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 795, 58-64 [10.1016/j.nima.2015.05.057].

Time-dependent electric field in Al/CdTe/Pt detectors

TURTURICI, Accursio Antonio;ABBENE, Leonardo
;
PRINCIPATO, Fabio
2015-01-01

Abstract

Abstract Al/CdTe/Pt detectors are very attractive devices for high-resolution X-ray spectroscopy, even though they suffer from bias-induced time instability (polarization). Polarization phenomena cause a progressive time-degradation of the spectroscopic performance of the detectors, due to hole trapping and detrapping from deep acceptor levels that directly control the electric field distribution. In this work we present experimental investigations on the electric field profile of planar Al/CdTe/Pt detectors by means of Pockels effect measurements. The time/temperature dependence of the electric field was investigated in a long time window (up to 10 h) and the correlation with the reverse current transients was also studied. Two energy levels (0.62 eV and 1.16 eV) of the deep hole traps were measured, in agreement with our previous results obtained through electrical and spectroscopic approaches.
2015
Turturici, A., Abbene, L., Franc, J., Grill, R., Dědič, V., Principato, F. (2015). Time-dependent electric field in Al/CdTe/Pt detectors. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 795, 58-64 [10.1016/j.nima.2015.05.057].
File in questo prodotto:
File Dimensione Formato  
Turturici_NIMA_2015.pdf

Solo gestori archvio

Tipologia: Versione Editoriale
Dimensione 2.14 MB
Formato Adobe PDF
2.14 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/149514
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 12
social impact