This paper reports the preliminary results obtained by Electron Paramagnetic Resonance (EPR) measurements on films of IRGANOX® 1076 phenols with and without low content (5% by weight) of gadolinium oxide (Gd2O3) exposed in the thermal column of the Triga Mark II reactor of LENA (Laboratorio Energia Nucleare Applicata) of Pavia (Italy). Thanks to their size, the phenolic films here presented are good devices for the dosimetry of beams with high dose gradient and which require accurate knowledge of the precise dose delivered. The dependence of EPR signal as function of neutron dose was investigated in the fluence range between 1011 cm−2 and 1014 cm−2. Linearity of EPR response was found and the signal was compared with that of commercial alanine films. Our analysis showed that gadolinium oxide (5% by weight) can enhance the thermal neutron sensitivity more than 18 times. Irradiated dosimetric films of phenolic compound exhibited EPR signal fading of about 4% after 10 days from irradiation.

Gallo, S., Panzeca, S., Longo, A., Altieri, S., Bantivolglio, A., Dondi, D., et al. (2015). Testing and linearity calibration of films of phenol compounds exposed to thermal neutron field for EPR dosimetry. APPLIED RADIATION AND ISOTOPES, 106, 129-133 [10.1016/j.apradiso.2015.07.043].

Testing and linearity calibration of films of phenol compounds exposed to thermal neutron field for EPR dosimetry

GALLO, Salvatore;PANZECA, Salvatore;LONGO, Anna;MARRALE, Maurizio
2015-01-01

Abstract

This paper reports the preliminary results obtained by Electron Paramagnetic Resonance (EPR) measurements on films of IRGANOX® 1076 phenols with and without low content (5% by weight) of gadolinium oxide (Gd2O3) exposed in the thermal column of the Triga Mark II reactor of LENA (Laboratorio Energia Nucleare Applicata) of Pavia (Italy). Thanks to their size, the phenolic films here presented are good devices for the dosimetry of beams with high dose gradient and which require accurate knowledge of the precise dose delivered. The dependence of EPR signal as function of neutron dose was investigated in the fluence range between 1011 cm−2 and 1014 cm−2. Linearity of EPR response was found and the signal was compared with that of commercial alanine films. Our analysis showed that gadolinium oxide (5% by weight) can enhance the thermal neutron sensitivity more than 18 times. Irradiated dosimetric films of phenolic compound exhibited EPR signal fading of about 4% after 10 days from irradiation.
2015
Settore FIS/01 - Fisica Sperimentale
Settore FIS/07 - Fisica Applicata(Beni Culturali, Ambientali, Biol.e Medicin)
Settore ING-IND/18 - Fisica Dei Reattori Nucleari
Settore ING-IND/20 - Misure E Strumentazione Nucleari
Gallo, S., Panzeca, S., Longo, A., Altieri, S., Bantivolglio, A., Dondi, D., et al. (2015). Testing and linearity calibration of films of phenol compounds exposed to thermal neutron field for EPR dosimetry. APPLIED RADIATION AND ISOTOPES, 106, 129-133 [10.1016/j.apradiso.2015.07.043].
File in questo prodotto:
File Dimensione Formato  
FILM_FENOLI_GALLO.pdf

Solo gestori archvio

Descrizione: Articolo per esteso
Dimensione 536.07 kB
Formato Adobe PDF
536.07 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/145072
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 16
social impact