This paper describes a portable apparatus to be utilized in mammographic X-ray spectroscopy under clinical conditions. The system, based on a CdTe solid-state detector, is able to directly measure mammographic X-ray tube spectra. Good system response to monoenergetic photons was measured using X-ray and γ-ray calibration sources (109Cd and 241Am). The measured molybdenum X-ray spectra, in agreement with simulated spectra, show the good spectral capability of the system also at high photon fluence rates, as typical of clinical mammography. Low tailing, no secondary X-ray escape and low pile-up distortions in the measured spectra indicate that this portable system is suitable for mammographic X-ray spectroscopy. © Società Italiana di Fisica.
LA MANNA A, L ABBENE, FFAUCI, GERARDI G, G RASO, S DEL SORDO, et al. (2006). Portable CdTe detection system for mammographic X-ray spectroscopy. LA RIVISTA DEL NUOVO CIMENTO DELLA SOCIETÀ ITALIANA DI FISICA, 29(3), 361-367 [10.1393/ncc/i2005-10237-7].
Portable CdTe detection system for mammographic X-ray spectroscopy
ABBENE, Leonardo;FAUCI, Francesco;GERARDI, Gaetano;RASO, Giuseppe;
2006-01-01
Abstract
This paper describes a portable apparatus to be utilized in mammographic X-ray spectroscopy under clinical conditions. The system, based on a CdTe solid-state detector, is able to directly measure mammographic X-ray tube spectra. Good system response to monoenergetic photons was measured using X-ray and γ-ray calibration sources (109Cd and 241Am). The measured molybdenum X-ray spectra, in agreement with simulated spectra, show the good spectral capability of the system also at high photon fluence rates, as typical of clinical mammography. Low tailing, no secondary X-ray escape and low pile-up distortions in the measured spectra indicate that this portable system is suitable for mammographic X-ray spectroscopy. © Società Italiana di Fisica.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.