The technique which combines high resolution spectroscopy with imaging capability is a powerful tool to extract fundamental information in X-ray Astrophysics and Cosmology. TES (Transition Edge Sensors)-based microcalorimeters match at best the requirements for doing fine spectroscopy and imaging of both bright (high count rate) and faint (poor signal-to-noise ratio) sources. For this reason they are considered among the most promising detectors for the next high energy space missions and are being developed for use on the focal plane of the IXO (International X-ray Observatory) mission. In order to achieve the required signal-to-noise ratio for faint or diffuse sources it is necessary to reduce the particle-induced background by almost two orders of magnitude. This reduction can only be achieved by adopting an active anticoincidence technique. In this paper, we will present a novel anticoincidence detector based on a TES sensor developed for the IXO mission. The pulse duration and the large area of the IXO TESarray (XMS X-ray Microcalorimeter Spectrometer) require a proper design of the anticoincidence detector. It has to cover the full XMS area, yet delivering a fast response. We have therefore chosen to develop it in a four-pixel design. Experimental results from the large-area pixel prototypes will be discussed, also including design considerations.

Macculi, C., Colasanti, L., Lotti, S., Natalucci, L., Piro, L., Bagliani, D., et al. (2010). The TES-based cryogenic anticoincidence detector for IXO: first results from large area prototypes. In SPIE - Space Telescopes and Instrumentation 2010: Ultraviolet to Gamma Ray (pp.77323Y-1-77323Y-11) [10.1117/12.856399].

The TES-based cryogenic anticoincidence detector for IXO: first results from large area prototypes

BARBERA, Marco;
2010-01-01

Abstract

The technique which combines high resolution spectroscopy with imaging capability is a powerful tool to extract fundamental information in X-ray Astrophysics and Cosmology. TES (Transition Edge Sensors)-based microcalorimeters match at best the requirements for doing fine spectroscopy and imaging of both bright (high count rate) and faint (poor signal-to-noise ratio) sources. For this reason they are considered among the most promising detectors for the next high energy space missions and are being developed for use on the focal plane of the IXO (International X-ray Observatory) mission. In order to achieve the required signal-to-noise ratio for faint or diffuse sources it is necessary to reduce the particle-induced background by almost two orders of magnitude. This reduction can only be achieved by adopting an active anticoincidence technique. In this paper, we will present a novel anticoincidence detector based on a TES sensor developed for the IXO mission. The pulse duration and the large area of the IXO TESarray (XMS X-ray Microcalorimeter Spectrometer) require a proper design of the anticoincidence detector. It has to cover the full XMS area, yet delivering a fast response. We have therefore chosen to develop it in a four-pixel design. Experimental results from the large-area pixel prototypes will be discussed, also including design considerations.
Settore FIS/05 - Astronomia E Astrofisica
28-giu-2010
SPIE - Space Telescopes and Instrumentation 2010: Ultraviolet to Gamma Ray
San Diego, California, USA
28 Giugno 2010
2010
11
A stampa
Macculi, C., Colasanti, L., Lotti, S., Natalucci, L., Piro, L., Bagliani, D., et al. (2010). The TES-based cryogenic anticoincidence detector for IXO: first results from large area prototypes. In SPIE - Space Telescopes and Instrumentation 2010: Ultraviolet to Gamma Ray (pp.77323Y-1-77323Y-11) [10.1117/12.856399].
Proceedings (atti dei congressi)
Macculi, C.; Colasanti, L.; Lotti, S.; Natalucci, L.; Piro, L.; Bagliani, D.; Brunetto, F.; Ferrari, L.; Gatti, F.; Torrioli, G.; Bastia, P.; Bonati, A.; Barbera, M.; La Rosa, G.; Mineo, T.; Perinati, E.
File in questo prodotto:
File Dimensione Formato  
2010SPIE_7732_3Y_Macculi_et_al.pdf

Solo gestori archvio

Dimensione 794.04 kB
Formato Adobe PDF
794.04 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/10447/56800
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 0
social impact