We report on the performance of our NTD-Ge microcalorimeters. To date, the spectral resolution for x-ray and gamma-ray lines from radioactive sources and laboratory plasmas is 4.8 eV in the entire 1 - 6 keV band and 52 eV at 60 keV. Technical details responsible for this performance are presented as well as an innovative electro-thermal approach for enhancing count-rate capability.
Silver, E., Bandler, S., Schnopper, H., Murray, S., Madden, N., Landis, D., et al. (2002). X-ray and gamma-ray astronomy with NTD germanium-based microcalorimeters. In AIP Conference Proceedings (pp.555-558) [10.1063/1.1457708].
X-ray and gamma-ray astronomy with NTD germanium-based microcalorimeters
BARBERA, Marco
2002-01-01
Abstract
We report on the performance of our NTD-Ge microcalorimeters. To date, the spectral resolution for x-ray and gamma-ray lines from radioactive sources and laboratory plasmas is 4.8 eV in the entire 1 - 6 keV band and 52 eV at 60 keV. Technical details responsible for this performance are presented as well as an innovative electro-thermal approach for enhancing count-rate capability.File in questo prodotto: